Heads for Measurement of Peak Power
- Laser Peak Power and Laser Pulse Shape Measurements
- Wide choice of Heads cover 100nW-100W and beams up to 60mm diam.
- Absolute Calibration Accuracy of ±2.5%
- Response times down to 1ns (10-90%)
- Heads accept parallel laser beams or wide angle light from an optical fibre
- Optional Digital Readout of peak power
Types of Laser Measured
All our Peak Power Measurement Heads are aimed at measuring the peak power of high repetition rate pulsed lasers. For semiconductor lasers, these might typically have a peak power of about 10W in 50ns but there are options for shorter pulses and higher or lower powers.
Peak Power Measurement Heads
The Peak Power Measurement Heads are designed to measure high repetition rate pulsed lasers, typically emitting about 10W in 50ns. The Heads for higher powers operate like integrating spheres and have a wide acceptance angle. They will even measure accurately the diverging light from the end of an optical fibre. The response times are close 1ns rise time.
The Low Power Heads have a lens to collect the light onto the photo diode. This allows the measurement of the very low powers returning after being scattered from a distant target (usually less than a microwatt). Although the lens reduces the acceptance angle to just a few degrees, this is easy to achieve in practice.
Signal Acquisition.
All the Heads are intended for use with either a fast 50 Ohm ’Scope or our Peak Power Module (see Specification at end of this web page).
Calibration.
All the Heads have built in Calibration facilities which we use to provide an accurate link between the power in a short high power pulse and the UK National Standards of CW power at about 1mW. The final absolute accuracy in the laser peak power measurement is typically better than ±2.5%. A sheet discussing the principles and accuracy of the calibration is available on request.
Power Supplies
The standard Heads for the higher powers (LD, SD) can be operated with a simple battery pack giving 54V bias for the photo-diodes. Thw others have internal amplifiers abd must be powered by the Power Supply Unit which also accepts the optional Signal Analysis Modules.
Diffuse Receiving Head (SD)
Lens Receiving Head (SL)
Peak Power Module
The Peak Power Module (PPM/1) captures the peak value of the signal from one of the Heads. This it does by a successive approximation process and hence is only suitable for a laser with PRF above about 20Hz. Its limited dynamic range means that in many cases the measurement is better done by a modern digitising 'Scope. However, it can still be useful when the pulses from the laser are slightly inconsistent because it has modes to display either the average peak power or the maximum peak power or the minimum peak power. It is also easier to use for Quality Assurance checks at a fixed point when the reduced dynamic range is not a problem.
It is completely automatic in action - two LEDs light up when the signals are being correctly processed.
The Display on the Module is Digital. There is also a steady output (for use with an external DVM) and a fast 50Ω output to allow monitoring of the signal on a'Scope.
Pulse Duration: 6ns to 2.5µs FWHM (less than 3% variation in readings)
Pulse Rep Rate: at PRFs up to 300 kHz every pulse is sampled. Above this frequency the pulses are sampled every 3µs. Min PRF for full accuracy is 20Hz.
Input Impedance: 50Ω (BNC on front of Module).
Sensitivity: An input of 200mV gives Full Scale Reading (2.000) on the DPM in the Module. The minimum input for correct operation is 10mV. The decimal point will be placed as appropriate to match the power range of the Head, reading in Watts. Other values of the sensitivity are possible.
Operational Modes: To allow for slight inconsistencies in the pulses from the laser, there is a choice of displaying the mean peak power or the maximum peak power, or the minimum peak power in the stream of pulses. The ‘max’ and ‘min’ modes find the 80 and 20 percentile points. Thus in ‘max’ setting the unit displays the value which has 80% of all pulses below it, while in ‘min’ setting it displays the value which has only 20% of pulses below it (80% above).
Meter Display: 3.5 Digit DVM
Fast Signal o/p: 1.5 ns rise time pulse from 50Ω. Amplitude is 1/2 input pulse.
Trigger Output: 0.3V for 2.5µs from reverse terminated 50Ω BNC. This is useful for 'scopes or External DVM
Display Time: Signal decays in about 30 sees if input pulses stop.
Signal Output: 2.0V from 1kΩ corresponds to FSD on the DVM.
Mean Power Module
Input: The input is via an internal link from the Peak Power Module.
Sensitivity: The sensitivity will be arranged to match the peak powers pulse widths and repetition rates of the pulsed input from the Head. Signal Output: 2.0V from 1 kΩ corresponds to FSD on the DVM.
Ambient Light: Because of the wide acceptance angle of the Head, ambient light may give an offset signal when measuring mean power. An offset knob allows this ambient light to be backed off.
Meter Display: 3.5 Digit DVM.
IEEE Interface
For both Signal Analysis Modules an IEEE-488 option is available for easy data collection by computer.